R3765/67G Series Network Analyzer Operation Manual
5.8 GO/NG Measurement Using Limit Line Function
5-48
The display on the screen is as follows.
Figure 5-41 Setting of Segment 0
Set segment 1, segment 2, segment 3 and segment 4 in the same way.
SEGMENT 1 :
{ADD SEGMENT}
{STIMULUS VALUE}
[8]
[4]
[0]
[MHz]
{UPPER LIMIT}
[-]
[4]
[0]
[X1]
{LOWER LIMIT}
[-]
[6]
[5]
[X1]
{Return}
SEGMENT 2 :
{ADD SEGMENT}
{STIMULUS VALUE}
[8]
[6]
[5]
[MHz]
{UPPER LIMIT}
[0]
[X1]
{LOWER LIMIT}
[-]
[2]
[0]
[X1]
{Return}