2) Scale the steps in accordance with the linear theUUT for the default.
3) Every point value and tolerance can be edited.
◆
Resolution.
Given: 1, 0.1, 0.01, 0.001.
◆
Cycles: 1~3.
◆
Temperature coefficient: used to correct the value where
applicable.
3. Run
◆
Input the information before running including the date
(the default is the system date), the ambient temperature
(the default is 25
℃
and the range is 0~50
℃
) and the
relative humidity (the default is 40%RH and the range is
0~100%RH).
◆
Stroke mode: One Way or Two Way.
◆
Wait for the displayed pressure to stabilize.
◆
Input the value from the UUT. After inputting the value for the UUT, the standard value will be recorded from the
calibrator.
◆
After completing all the set points a report is automatically generated.
5.3.3 Aneroid barometer
1. Connect
◆
Connect the calibrator to the aneroid barometer as shown in Figure 5-3.
2. New and Run
Creating and running the task are similar with calibrating a mercurial barometer (Section 5.3.2).
Figure 5-3 Calibrating an aneroid barometer
53