3.5.5 Calibrating pressure switches
1. Connect
◆
Connect the calibrator to pressure switch as shown in Figure 3-23.
2. New
◆
Task name
1) Enter an alphanumeric name, and press numerical keypad to
input special codes.
2) The name cannot be the same as the existing task.
3) The characters of the input test name cannot exceed 13 spaces.
◆
Pressure type: Gauge, Absolute or Differential.
◆
Range: Pressure range of the switch.
◆
Dead band range: is the minimum and maximum of the dead band,
Dead band is the pressure difference between the change-of-states
(i.e. OPEN-CLOSE or CLOSE-OPEN).
◆
Set point: The pressure which the switch will change state.
◆
Tolerance: The allowable variation from the set point pressure.
◆
Trip type: High or Low. This is the direction to which the
change-of-state should happen. If the trip type is low, this means the
change-of-state happens when the pressure
is falling. If the trip type is high, this means the change-of-state happens when the pressure is rising.
◆
Cycles: 1~3.
Figure 3-23 Switch Terminology