Entering any value with the numeric key pad will show the corresponding values in all units.
3.4.7.4 Help
The calibrator provides images and diagrams showing basic connections and typical applications.
◆
Select the Help in the setup screen.
◆
Press / to view.
3.4.8 Information
◆
Sys Info: Model, serial number, software rev, hardware rev, and manufacturing date.
◆
Atmos.: Provides the atmospheric pressure.
◆
Hi and low Range: Model, range, overload, manufacturing date, serial number, form (pressure type), accuracy, and
software rev.
◆
P. Module: If an external pressure module is connected, the model, range, overload, and manufacturing date
information is displayed.
3.4.9 System Calibration
◆
The calibrator should be powered to warm up 30 minutes prior to starting the calibration.
◆
To ensure optimum instrument operation, the current/voltage measure, the current source and the internal dual
pressure range should be calibrated at regular intervals.
◆
The calibrator also provides a way to calibrate the external ADT160 pressure module.
◆
The calibrator can also calibrate inlet pressure and atmosphere pressure.
◆
Making errors when calibrating will affect the precision and possibly even cause abnormalities.
◆
The calibration data can be restored to the factory default, and the calibration date changes to “----/--/--”.
◆
Recommend the use of appropriately accurate primary standards when calibrating this instrument.
◆
Enter the system calibration in the Setup screen and select the function.
◆
The password is 811.