Task name
1) Enter an alphanumeric name, and press numerical keypad to input special codes.
2) The name cannot be the same as an existing task.
3) The characters of the task name cannot exceed 13 spaces.
◆
Range of UUT
1) Cannot exceed the output range of the calibrator.
2) The high and low limit cannot be equal.
3) The calibrator will automatically select the range for higher precision.
4) The calibrator will use dual range if needed.
5) Over pressuring can result in damage to the UUT.
◆
Pressure type: gauge, absolute or differential.
◆
Accuracy
1) Preselected accuracies are available.
2) Custom X.XX*: When the custom range is selected, press F2, and the field can be edited to the desired accuracy
from (0.01 to 100)%FS.
◆
Calibration points
1) The range is 2 to 17.
2) Scale will default to linear steps over the chosen range of the UUT.
3) Each test point can be edited individually.
4) Setting a point to less than -0.95bar will be adjusted to -0.95bar.
◆
Resolution
1) To format the indicator of the UUT.