Stroke mode: One Way or Two Way.
◆
Wait for the displayed pressure to stabilize. Input the value from the UUT. After inputting the value for the UUT, the
standard value will be recorded from the calibrator.
◆
After completing all the set points a report is automatically generated.
5.3.2 Mercury barometer
1. Connect
◆
Connect the calibrator to the mercury barometer as shown in Figure 5-2.
◆
Connect the REF/FLT port to the low pressure port of the differential pressure gauge.
2. New
◆
Task name:
1) Enter an alphanumeric name, and press
to input special codes.
2) The name cannot be the same as the existing task.
3) The characters of the input task name cannot
exceed 12 spaces.
◆
Range of UUT:
1) Range: 0.1bar~1.2bar.
2) Cannot exceed the output range of the calibrator.
3) The high and low limit cannot be equal.
4) Over pressuring will result in damage to the UUT.
◆
Calibration points:
1) The range is 2-17.
Figure 5-2 Calibrating a mercurial barometer
52