12
④
Status bar: Run mode (not visible in the home screen), snapshot, HART online, pressure module online, 24V
power status, battery level.
⑤
Triple display: If the triple display is enabled, then three windows will be shown when the pressure module is connected.
⑥
The pressure output mode: Press / to change to standby mode, exit the control mode, or to vent the
calibrator.
⑦
Pressure units menu: Up to 11 selectable pressure units.
3.1.2 Pressure output
1. Select High/Low Pressure Range in source mode.
◆
The calibrator must be vented or in standby mode.
◆
The calibrator will automatically vent when switching ranges.
2. Connection: Complete the pneumatic connection as shown in
Figure 3-3 or Figure 3-4.
3. Entering a value
◆
Use the numeric keypad to enter the value, and press
/ to confirm.
◆
If there is no operation within 10 seconds, the calibrator will
automatically cancel the entry.
◆
The value must meet the control range of the calibrator.
◆
The maximum negative pressure is -13.5 psi (-0.95 bar).
But this is dependent on the local atmospheric pressure.
◆
The value cannot exceed the control range of the CONTROL SETUP,
or the calibrator will prompt the user to reset the limits.
、 、 、
Figure 3-3 Gauge pressure source
Figure 3-4 Differential pressure source
REF/FLT