5 Special Application for Barometric Pressure
The ADT761-BP is a unique calibrator and different from all the other ADT761 series calibrators.
5.1 ADT761 Basic differences
◆
Single range: 10kPa ~120kPa, absolute sensor type.
◆
Home screen: Does not support the triple display and press to switch to display mode between single window and
two windows.
◆
Zero: This function is not available.
5.2 Gas head correction
◆
If enabled, the calibrator will display the correct result automatically according to the setting of parameters.
◆
Set the Height value in accordance with the Metric or English Units.
◆
Set the Density value in accordance with the Medium (i.e.Nitrogen, Dry Air, etc)
◆
Select standard gravity 980.665 cm/s or the local gravity based on
the calibration location.
5.3 Specific ADT761-BP applications
5.3.1 Digital barometer
1. Connect
◆
Connect the calibrator to the digital barometer as
shown in Figure 5-1.
2. New
◆
Task name:
1) Enter an alphanumeric name, and press
to input special codes.
Figure 5-1 Calibrating a digital barometer