21
◆
Ensure a proper connection (see connections above). From the home screen, press to start the HART function.
◆
The calibrator will poll address 0. If no connection is made, it will scan from address 1 to 15.
◆
Once the calibrator establishes communication with the transmitter, the transmitter will be added into the HART
device list.
◆
Press / to stop the polling.
◆
Press to restart polling.
3. Online/offline
◆
The HART functions are shown in the Measure window of the Home screen and the HART icon will display in
the status bar after connection to the transmitter is established.
◆
Press / to disconnect the HART instrument and the HART icon will disappear.
◆
If HART communication is terminated, the HART function will abort and the calibrator will prompt to poll again.
◆
If HART communication is terminated in use, the HART function will abort.
3.2.2 Process variables
◆
Display HART-PV, HART-AO, HART-%, Loop mA.
◆
Press / to select the process variable to be the main value displayed in the MEASURE window.
◆
Select the process variable in the MEASURE screen before running a HART transmitter calibration task.