A graph is shown depicting each step of the test.
◆
Press to vent and end the test.
◆
The set point with the internal ranges cannot exceed the control range of the Control Setup, or the calibrator will
prompt the user to reset the limits.
3.4.7.2 Vent
This operation will clean out impurities from the calibrator by pressurizing the calibrator and venting the system.
1. Connect
◆
Plug the Outlet port with the provided plug.
◆
Please gather and properly dispose of any pollutants during venting as shown in Figure 3.20.
◆
The vent port should not face the operator.
2. Run
◆
Select vent under the tools icon in the setup screen.
◆
The 761 will automatically switch to the high range.
◆
Press / to pressurize to the venting pressure (Set Point).
◆
When the pressure reaches the venting pressure, the vent valve will open to release the system gas and other
pollutants.
◆
Venting can be performed repeatedly.
◆
The set point cannot exceed the control range of the Control Setup, or the calibrator will prompt the user to reset the
limits.
3.4.7.3 Unit conversion
◆
Select converter under the tools icon in the setup screen.
◆
Eleven pressure units are show in the conversion chart.