1) (Output value-Set value) <= control stability setting.
2) Control stability is achieved within the stable delay period.
◆
Stable beep: the stable prompt sound.
4. Slew rate
◆
Pressurize at a fast/medium/slow speed.
◆
The default is fast but may be changed to better determine the precise trip value when testing a pressure switch.
5. Switch damping
◆
Averaging time period before the switch trips. The range of the time is (0.1~99) seconds and the default is 0.1.
◆
Set it according to the switch specifications, too long of a dampening time may cause an error in the trip point.
6. Auto Zero
◆
If enabled, the calibrator will automatically zero pressure value when it is in vent state and record the offset.
◆
If Absolute pressure type selected, the Auto Zero will be disabled automatically.
7. Vent P. (Vent Pressure)
◆
When the real pressure is less than the Vent Pressure setting and when controlling the zero pressure, the calibrator
will automatically vent.
◆
Accepted range of Vent P.: (0~105) %FS.
3.4.2 24V power
◆
Short current protected, then the symbol indicating 24V status on the status bar will flicker.
◆
The DC24V power will stay activated when using the current Source or the HART function with internal loop power and
will return to the original status after the loop power is changed.
3.4.3 Display setup
1. Resolution
29