Resolution: the value is 1, 0.1, 0.01, 0.001, and the default is 0.1.
◆
Cycles: 1~3.
3. Run
◆
Input the information before running including the date (the default is the system date), the ambient temperature (the
default is 25
℃
and the range is 0~50
℃
) and the relative humidity (the default is 40%RH and the range is
0~100%RH).
◆
Stroke mode: One Way or Round Trip.
◆
Wait for the displayed pressure to stabilize.
◆
Input the value from the UUT. After inputting the value for the UUT, the standard value will be recorded from the
calibrator.
◆
After completing all the set points a report is automatically generated.
5.3.5 Pressure transmitter
As shown in section 3.5.3 and section 3.5.4.
5.3.6 Pressure switch
As shown in section 3.5.5.