Select the High or Low range in SOURCE mode.
◆
Select the Current or Voltage in MEASURE mode.
◆
Activate 24V power.
2. Connect
◆
Complete the pneumatic connection as
shown in Figure 3-17.
◆
For low range differential sensors, connect the
REF/FLT port to the reference port of the gauge
as shown in Figure3-4.
3. Pressurize manually or by the step function
(Section 3.1.11).
4. Record the measure value manually or by using
snapshot (Section 3.1.12) after the pressure has stabilized.
3.3.3 Calibrating HART transmitters
1. Prepare
◆
Select the High or Low range in SOURCE mode.
◆
Select the current in MEASURE mode.
◆
Activate 24V power.
2. Connect
◆
Complete the pneumatic and the electrical connections as shown in Figure 3-13 and Figure 3-17, when using
internal series resistor and 24V loop power.
Figure 3-17 Calibrating a pressure transmitter
25
/FLT
Содержание ADT761 Series
Страница 1: ...ADT761 Series Automated Pressure Calibrator...
Страница 14: ...7 2 1 4 Keypad Table 2 3 Key functions...