to descend, the delay time will count down to zero, then the task turns to the descending points.
2) Manual: Press / to continue after the pressure is stabilized.
◆
Start and wait for the displayed pressure to stabilize, and then record the measured value. Complete this step and
continue.
◆
Source the current of the set points until the end and generate a report automatically.
◆
Recall the previous steps: press / to pause and the calibrator will switch to the Standby mode, and
then press / to recalibrate this set point.
4. Troubleshooting
4.1 Device information
◆
Select the device information in the setup screen.
◆
View the software revision, factory date, serial number and so on.
◆
View the information of the high / low range inside.
◆
View the current atmospheric pressure.
4.2 Prompt
The display shows a screen prompt warning as follows:
Error message
Description
As the pressure type of internal pressure sensor is gauge the calibration should
be matched with gauge pressure
Under the condition of system leakage the self tuning function will lead next
steps to errors So the users must check if the plugs is well connected and try
again
The calibration will be automatically matched to gauge
pressure to match the internal sensor type
Unable to control Please check the system for leaks
47
Содержание ADT761 Series
Страница 1: ...ADT761 Series Automated Pressure Calibrator...
Страница 14: ...7 2 1 4 Keypad Table 2 3 Key functions...