3.4.5 Language
◆
Standard: English, Spanish, Simplified Chinese, German, French, Italian, Portuguese.
◆
Optional: Traditional Chinese, Japanese or Russian available upon request.
3.4.6 Factory default
◆
Restore instrument settings to factory default including the Control Setup, the Display Setup and the RS232 Setup. It
also deletes all the snapshots and task files.
◆
The password is 811.
◆
Please note that all new settings will be erased.
3.4.7 Tools
3.4.7.1 Leak test
1. Range Selection
◆
Hi. Range/Lo. Range: Select the internal high/low range for pressurizing to the leak test point automatically with the
calibrator.
◆
P. Module: pressurize with an external pressure source and measure the real pressure with an external pressure
module.
2. Setup
◆
Select the leak test under the tools icon in the setup screen.
◆
Leak test with the internal ranges or the external pressure module.
1) The set point and the wait time do not have to be set with the external pressure module.
2)The pressure unit can be selected. The default unit is kPa. (Section 3.1.4).
◆
The pressure unit will return to the previous selected unit for pressurization (Section 3.1.4).
◆
Set point
1) Do not exceed the range of the UUT.
2)With the numerical key pad enter the leak test point.
31