2101510 Rev. AG
Page 6–7
Additionally, component high/low concentration, component
peak not found, and component RF limit exceeded alarms
are available but disabled. These alarms may be enabled by
the user, but are not included here for the purposes of
troubleshooting. See the PCCU32 help files for more
information.
Table 6–1 NGC8200 Alarms
Description
Enable Type
Severity
Pressure Regulator 1
Yes
GT
Fault
Pressure Regulator 2
Yes
GT
Fault
Sample Pressure
Yes
GT
Fault
Oven Temperature Error
Yes
GT
System Fault
No Stream Valve Selected
Yes
GT
System Fault
Digital-Analog Bd Comm Error
Yes
GT
System Fault
Calculation Error
Yes
GT
Fault
Calibration Un-Normalized Total
Yes
GT
Fault
Stream Sequence Error
Yes
GT
Fault
Calibration CV Percent Error
Yes
GT
Fault
RF Pct Error
Yes
GT
Fault
Analog Bd Ambient Temp
Yes
GT
Warning
Analog Power Supply
Yes
GT
Warning
Low Carrier Gas Bottle (DI1)
Yes
LT
Warning
Low Cal Gas Bottle (DI2)
Yes
LT
Warning
GCM Chrom Process
Yes
GT
System Fault
Bad Bead
Yes
GT
Fault
No Pilot Valve Change Detected
Yes
GT
Fault
Sample Flow Detect
Yes
GT
Fault
CPU Loading
Yes
GT
Warning
System Memory Available
Yes
LT
Warning
Ram File Available
Yes
LT
Warning
Flash File Available
Yes
LT
Warning
Missing Peak-Cal Not Used
Yes
GT
Warning
Stream Un-Normalized Total
Yes
GT
Warning
6.3.1
Operators
•
GT = Greater Than
•
LT = Less Than
•
And = Including
•
Or = Instead of
•
GE = Greater Than or Equal To
•
LE = Less Than or Equal To
•
NAND = And Not
•
Nor = Not Or
Содержание NGC8206
Страница 1: ...2101510 rev AG NGC8206 Chromatograph User s Manual ...
Страница 14: ...xii Figure 6 3 AC Charger Power Supply Wiring 6 28 Figure 6 4 Communication Troubleshooting Flowchart 6 30 ...
Страница 27: ...Page 2 8 2101510 Rev AG Figure 2 4 NGC8206 Enclosure Figure 2 5 NGC8206 Enclosure Left Side ...
Страница 60: ...2101510 Rev AG Page 2 41 hex socket set screw on cap Figure 2 32 Explosion Proof AC Power Supply ...