![background image](http://html2.mh-extra.com/html/zeiss/opmi-lumera-700/opmi-lumera-700_user-manual_3357350090.webp)
System Overview
OPMI LUMERA® 700
Version 7.3
Page 90
G-30-1673-en
Connector panel on the floor stand
1
Strain relief device
for securing the power cord and video connection cable
2
Potential equalization connector
3
Cable connection for the foot control panel
4
Remote connector
for controlling external devices with a maximum breaking capacity of 24V/
0.5A.
5
Service interface
for exporting error messages for Carl Zeiss Service.
.
6
RJ45 Ethernet connector (option)
For networking OPMI LUMERA 700 and the CALLISTO eye OR manage-
ment system.
7
Sliding switch for rated voltage
The voltage shown here must correspond to the rated line voltage pro-
vided on the site of installation. You can adjust the sliding switch using a
suitable tool.
8
Power inlet socket
9
Power outlet socket
for medical devices of max. 500 VA
With the direct network connection between OPMI LUMERA 700 and
CALLISTO eye (optional, the two devices must first be interconnected using a
LAN cable. Afterwards, OPMI LUMERA 700 must be switched on and then
CALLISTO eye.
NOTE
Device fault!
• No USB hub with a power supply may be plugged into the service inter-
face.
CAUTION
Connection to data networks!
Use the LAN port only in combination with "galvanic network isolation for
medical electrical equipment" in accordance with DIN EN 60601-1.
The data network connector must have adequate protection against contact
at the end of cable to the separator (network side), e.g., made of plastic ma-
terial.
The cable and connector of the network connection must at least comply
with Cat-5e EIA/TIA-568A-5, i.e., the more recent Class D values from ISO
11801:2002 or EN 50173-1:2002.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...