![background image](http://html2.mh-extra.com/html/zeiss/opmi-lumera-700/opmi-lumera-700_user-manual_3357350146.webp)
Preparations for use
OPMI LUMERA® 700
Version 7.3
Page 146
G-30-1673-en
Connecting Invertertube E and RESIGHT 700 (optional)
to the surgical microscope
The two Invertertube E units are mounted (see page 142).
• Plug the connecting cable on Invertertube E into the Invertertube connec-
tion (2) on the surgical microscope.
• Plug the connecting cable on Invertertube E into the Invertertube connec-
tion (1) on the assistant's microscope.
The motorized image inversion function of of the two Invertertube E
units can now be controlled via the system.
• Plug the RESIGHT 700 connecting cable (above figure) into the
RESIGHT 700 connection (3) of the surgical microscope.
The internal focus of the RESIGHT 700 fundus viewing system
can now be controlled via the system. The internal focus ensures that
the surgical microscope no longer needs to be vertically moved. When
working on the anterior segment of the eye, you can therefore move
the fundus viewing system out of the beam path without causing blur-
ring in the surgical microscope.
If both devices are connected simultaneously, they are synchronized so
that the field of view is always displayed in the correct orientation. This
also includes the image inversion function of an integrated SD 3CCD or HD
3CCD camera.
Fixing the connecting cable with a cable clip
To make sure that the connecting cable of the Invertertube E on the as-
sistant's microscope does not make contact with the sterile operating caps on
the surgical microscope, you can fix the connecting cable using a cable clip:
• Push the dark gray cap out of the cable clip (Fig. B).
• Form the connecting cable of the Invertertube E into a loop (Fig. C).
• Pass this cable loop through the cable clip (Fig. D).
• Pull the cable loop out of the cable clip until the
connecting cable runs straight up and no longer hinders you (Fig. D).
• Push the dark gray cap back into the cable clip to fix the loop (Fig. E and F).
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...