Safety Measures
OPMI LUMERA® 700
Version 7.3
Page 14
G-30-1673-en
• Check the surgical microscope for sufficient freedom of movement. The
device itself or accessories may be damaged if conflicting with each other.
• When mounting accessories (e.g. fundus viewing systems) on the system,
make sure that sufficient free space is available for focus positioning and
that the surgical microscope cannot touch the patient.
• Before using a fundus viewing system, make sure that the room for move-
ment is larger than the downward travel path of the microscope.
• In order to prevent an unexpected response by the device, check the user
settings of the software before every use.
• The assistant's microscope may swing out of position if you tilt the surgical
microscope to the horizontal viewing direction and the assistant's micro-
scope has not been locked. Adjust the assistant's microscope before use
and make sure it is properly locked in position.
• Always run cables in a manner that will not impede the user's movements.
• Before every use, switch off the light source for light guides that are not
being used and properly attach them on the designated bracket.
• Before every use, check the light source for damages in order to prevent
phototoxic injuries to the patient's eye.
• Before every use, check the filter settings in order to prevent phototoxic
injury to the patient's eye.
• Use the system in the sterile area. Make sure to use the appropriate sterile
accessories for this system.
• Never cover any ventilation openings, as this may e.g. cause the light
source of the system to overheat and fail.
• Never attempt to forcefully connect any electrical connectors (plugs, sock-
ets). If connection is not readily possible, check whether the plug fits the
socket. If any of the connectors are damaged, have Carl Zeiss Service re-
pair them.
• Before using the
wireless
, foot control panel FCP WL ensure its batteries
are fully charged. Inadequate power supply of the
wireless
foot control
panel may lead to malfunctions of the device.
• Go through the checklist in chapter "Operation" (see page 164).
More specific information about the floor stand:
• In order to avoid the device from rolling off inadvertently, use the locking
mechanism on the base of the suspension system and secure the device.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...