Version 7.3
G-30-1673-en
Page 15
OPMI LUMERA® 700
Safety Measures
During use
• In order to prevent phototoxic damages to the patient's eye, reduce the
brightness and the light exposure of the surgical area to an amount that
is absolutely necessary for the surgery.
• In order to prevent phototoxic damages to the patient's eye during sur-
gery, use a retina protection filter (blue barrier filter) and/or a retina pro-
tection device that are integrated into the light source.
• In order to prevent phototoxic damages to the patient's eye, adjust the ir-
• In order to prevent phototoxic damages, never look directly into the light
source, e.g. on the microscope or the light guide.
• Never leave a device unattended with the light source still switched on, as
excessive radiation exposure times may lead to retinal injury in the pa-
tient's eye.
• When operating the handgrips or the zoom knob on the assistant's micro-
scope, do not touch the non-sterile connecting cable to avoid infecting
the patient.
• If you have mounted the RESIGHT fundus viewing system on the underside
of the microscope and steeply tilt the microscope, the RESIGHT fundus
viewing system may unintentionally move in and injure the patient.
Remove the RESIGHT fundus viewing system before steeply tilting the mi-
croscope.
• Defective or unidentified accessories may lead to increased leakage cur-
rent on the system and injure the patient. Never connect any defective or
unidentified accessories and never touch the power outlet or video inter-
faces while in contact with the patient.
• If the xenon lamp is used beyond its maximum service life of 500 hours,
sudden failure may occur and interrupt operation. Once the xenon lamp's
maximum service life reaches 5 hours, a warning is displayed, asking you
to replace the relevant xenon lamp. Swing in the backup xenon lamp now
at the latest. If the maximum service life of the backup xenon lamp has
also been reached, replace the xenon lamp module of the Superlux Eye
light source and reset the remaining service hours counter to its original
value.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...