System Overview
OPMI LUMERA® 700
Version 7.3
Page 70
G-30-1673-en
Components of the XY coupling
Controls of the XY coupling
1
XY reset button with LED indicator
You can have the functions listed below reset to their initial value when
you press the XY reset button or move the suspension arm upward into
the standby position. The XY reset button can be configured via the
5.7" control panel to meet your specific needs (see )page 186).
– XY coupling reset
When you press the XY reset button, the XY coupling moves to its
center position.
– OPMI focus reset (always active)
– RESIGHT 700 focus reset (always active)
When you press the XY button, the focus of a connected fundus view-
ing system RESIGHT 700 moves to its center position.
The initial magnification level can be defined via the 5.7" control panel
(see page 222).
Status of the reset button
– LED off: the button has been pressed and the system has been reset
to the preselected initial status.
– LED on: a function has been activated and the system is no longer in
the initial status.
CAUTION
Risk of injury!
When pressing the XY reset button, the device may be put in focus and can
be lowered.
• Ensure the working distance between microscope and the patient is al-
ways larger than the moving range of the microscope.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...