Preparations for use
OPMI LUMERA® 700
Version 7.3
Page 142
G-30-1673-en
Mounting accessories on the surgical microscope
Replacing components and accessories
Replace components as follow:
• Put the suspension arm in horizontal position, pull out the safety
button (1) and turn it 180° to the right or left.
Move the suspension arm slightly upward and downward until the lock
snaps in. Now the suspension cannot swing upward due to missing mass.
• Loosen securing screw (6) by a few turns.
• Remove dust cover (2) and store it in a safe place.
CAUTION
Risk of injury!
The swivel arm may move downward when accessories are changed.
• Never exceed the maximum permissible weight capacity of the suspen-
• Re-adjust the balance of the suspension arm after every change of acces-
• Adjust the limit of downward travel of the suspension arm in such a way
CAUTION
Risk of injury!
Incorrectly mounted accessories can fall down.
• Do not replace accessories over the patient.
• Before every use and after re-equipping the system, make sure that the
accessories are securely locked in position. Make sure that securing
screws (6) and (10) have been firmly tightened!
• Mount accessories such as the slit illuminator VISULUX (10) or
RESIGHT (8) only in accordance with the instructions of use.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...