![background image](http://html2.mh-extra.com/html/zeiss/opmi-lumera-700/opmi-lumera-700_user-manual_3357350128.webp)
Preparations for use
OPMI LUMERA® 700
Version 7.3
Page 128
G-30-1673-en
Positioning and connecting the floor mount
Positioning the device in the OR
Positioning the ceiling mount with lift arm
1
Standby position
2
Working position
Proceed as follows to position the ceiling mount with lift arm:
Set working position
• To lower the lift arm, turn the lift switch (4) to the right and hold it in that
position until the lift arm has reached the desired working height.
• After releasing the lift switch (4), the lift arm is locked at its working
height.
• While pressing and holding the brake release button, use the hand
grips (3) to move the surgical microscope to the desired working
position (2).
Setting the standby position
• For mounted carrier arm systems please additionally note page 132.
• While pressing and holding the brake release button, use the hand
grips (3) to move the surgical microscope to the topmost position.
CAUTION
Risk of injury to the patient!
Making a motor-driven lift arm readjustment during the operation might in-
jure the patient.
• Always adjust the lift arm without patient before starting the operation.
• Do not readjust the lift arm as long as the surgical microscope is posi-
tioned above the patient.
• Do not use the motor-driven lift arm for focusing.
CAUTION
Injury to the patient!
If the lift arm is moved with locked suspension arm, the device may be dam-
aged and the patient may be injured by dropping parts of the cover.
• Never lock the suspension arm with the locking screw (5) while your are
moving the lift arm.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...