Indicator for the operating status of the LED light source
If the LED lights up amber, the LED light source is defective. In this case,
the light intensity is reduced to 50% and an error message is displayed on
the 5.7” control panel.
• Please contact Carl Zeiss Service.
11
Flap indicating the operating status of the halogen lamps
– When the flap is closed, the main lamp is operative.
– When the flap is open (yellow mark on the side), the main lamp has
failed and the backup lamp is in use.
12
"Manually turn on the backup halogen lamp" button
The halogen light source is equipped with an automatic lamp backup. If
the backup fails, pressing this button will enable you to switch on the
backup lamp manually.
13
"Pivot in xenon backup lamp" button
The Superlux Eye light source is equipped with a lamp module that con-
tains two xenon lamps. The second lamp is used as a backup lamp which
can be pivoted in manually if the first lamp malfunctions. (see page 244)
CAUTION
Lamp failure!
– The service life of the xenon lamp is limited to 500 hours. If the service
life of the lamp is exceeded, the xenon lamp may suddenly fail.
– With increasing age of the light source, the actual illumination intensity
delivered at the respective setting decreases.
• Please replace the xenon lamp in due time.
• Reset the residual service hour display on the 5.7" control panel to "0".
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...