Preparations for use
OPMI LUMERA® 700
Version 7.3
Page 148
G-30-1673-en
Connecting the second light guide to the accessories
Before using the system, check whether the second light guide is connected
to the accessory equipment. If this is not the case, proceed as follows to con-
nect the second light guide:
• Switch off the light source.
• Insert the appropriate end of the light guide into the light guide socket of
the accessory equipment as far as it will go.
Fig. 70: Connecting the light
guide
CAUTION
Risk of injury!
A second light guide, which freely hangs from the surgical microscope, may
cause retinal damage or slight burns to the patient.
• Always turn off the light source for the light guide that is not in use.
• Press the unused light guide into an open position of the cable holder (1)
in such a way that it points away from the patient.
• Remove the protective cap of the unused light guide. This will prevent
melting of the protective cap if the light source is accidentally turned on.
2
1
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...