Safety Measures
OPMI LUMERA® 700
Version 7.3
Page 16
G-30-1673-en
• If you change the lamp shortly after it has failed, the lamp will still be very
hot. Use heat-resistant protective gloves when replacing the lamp to pre-
vent burns.
• Malfunctions in the motor's electronic system may cause the main func-
tions to fail (XY movement, focus, zoom, light control system) and affect
other functions. Change to manual mode.
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the device stat-
ing it is out of order and contact our service representative.
• Do not pull at the light guide, power cord or any other connecting cables.
After every use
• Always use the master switch to turn off the device if it is not in use.
• Insufficient, incorrect or wrong cleaning or disinfection not complying
with this user manual can expose the patient or medical staff to a consid-
erable risk of infection.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...