OPMI LUMERA® 700
System Overview
IDIS configuration and operation
The following prerequisites must be met in order to use the assistance func-
tions of
CALLISTO eye:
To enable use of the K TRACK assistance function, OPMI LUMERA 700
must feature an integrated keratoscope.
The configuration and operation of IDIS and its assistance functions are de-
scribed in the CALLISTO eye operating manual (G-30-1708).
Fig. 24: Integrated Data
Injection System
NOTE
Poor surgical outcome!
If data errors occur or wrong optics parameters are input, the capsulorhexis
ring is injected in the eyepiece with an offset and the surgical outcome dete-
riorates.
• Check the entered optics parameters before surgery.
• Check whether the injected ring data is plausible with regard to position
and diameter.
NOTE
Defective IDIS!
When the IDIS is defective, assistance functions may fail and image errors
such as offset image data may occur.
• To prevent poor surgical results such as incompletely corrected astigma-
tism, disable IDIS and align the toric lens using the CALLISTO eye screen.
12
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...