Adjusting the surgical microscope
Adjusting the downward travel limit
When setting the stroke limits, proceed as follows:
• Loosen the adjustment screw (1) by a few counterclockwise turns.
• Press one of the release buttons for the magnetic brake on the surgical mi-
croscope and lower the surgical microscope to a point where focusing in
the operating field is still possible (depending on the focal distance of the
objective lens). Ensure that there is still sufficient safety distance to the op-
erating field.
• Turn the adjustment screw (1) clockwise as far as it will go.
• Lower the surgical microscope again to the lowest limit stop and check
the safety distance.
Fig. 72: Setting
downward travel
CAUTION
Risk of injury!
An improperly balanced systems will move upward or downward.
• Limit the stroke of the suspension arm to prevent any contact with the
patient if the surgical microscope is lowered accidentally.
1
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...