Preparations for use
OPMI LUMERA® 700
Version 7.3
Page 114
G-30-1673-en
Connecting the system to the power supply
• Make sure that power switch (1) is OFF.
• Insert the supplied power cord in power inlet socket (4) on the suspension
system.
• Only connect the power cord to wall outlets which are provided with a
properly connected protective ground conductor.
• Secure the power cord using the strain relief device (see page 118).
• If required, connect the suspension system to the potential equalization
contact in the OR using potential equalization pin (3).
NOTE
Tripping hazard!
Cables inappropriately routed on the floor may constitute a tripping hazard.
• Always route cables in a manner that will not obstruct the workflow.
NOTE
Risk of damaging the device!
Prior to leaving the factory, the device is adjusted to the rated voltage of the
respective export country. The rated voltage displayed on the sliding
switch (5), must correspond with the rated voltage of the place of installa-
tion.
• If the rated voltage is not correctly displayed, use an appropriate tool to
set the sliding switch (5) accordingly.
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...