OPMI LUMERA® 700
Safety Measures
– in explosion-hazardous areas,
– do not operate the device close to flammable anesthetics or volatile
solvents, such as alcohol, ether/benzene or similar. Stay away at least
25 cm.
• Do not station or use the instrument in damp rooms. Do not expose the
instrument to water splashes, dripping water or sprayed water.
• Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
• Never look at the sun through the binocular tube, the objective lens or an
eyepiece.
• The control unit features a potential equalization connector. This connec-
tor can be used for placing other active devices at the same ground po-
tential or for redundant grounding to protective earth.
• Do not use multiple sockets!
• Never open the device! The device contains freely accessible live compo-
nents. If you remove the housing, you run the risk of electric shock.
• Do not touch the system if your body is electrostatically charged and the
system is not grounded.
• Before connecting the system to line power, make sure that the power
network is free from defects that may lead to dangerous voltage on the
power plug.
• Please note the information on EMC (electromagnetic compatibility) in the
chapter "System data" on page 285.
• With increasing age of the light source, the actual illumination intensity
delivered at the respective setting decreases (this is a normal property of
the system).
• The user (IT officer) is responsible for ensuring that no viruses are trans-
ferred to the system via the network connection.
• It is the user's responsibility to ensure that the USB storage media used for
data communication are free from viruses.
• Should the device be transported over longer distances (e.g., moving, re-
turn for repairs etc.), only use the original packaging or special return
packages. Please contact your dealer or the Carl Zeiss Service department.
• In order to prevent a decrease in the device's safety due to age, wear and
tear, etc., the user must have the device checked for its safe operation (re-
fer to Safety Checks in Chapter "Care and maintenance").
More specific information about the floor stand:
Summary of Contents for opmi lumera 700
Page 1: ...Carl Zeiss OPMI LUMERA 700 User manual G 30 1673 en Version 7 3 2012 05 02...
Page 4: ...OPMILUMERA 700 Version 7 3 Page 4 G 30 1673 en...
Page 46: ...Safety Measures OPMI LUMERA 700 Version 7 3 Page 46 G 30 1673 en...
Page 61: ...Version 7 3 G 30 1673 en Page 61 OPMI LUMERA 700 System Overview Fig 23 Overhead display 11...
Page 106: ...System Overview OPMI LUMERA 700 Version 7 3 Page 106 G 30 1673 en...
Page 109: ...Version 7 3 G 30 1673 en Page 109 OPMI LUMERA 700 Preparations for use...
Page 113: ...Version 7 3 G 30 1673 en Page 113 OPMI LUMERA 700 Preparations for use Fig 48 OR scene 1 2...