6 Care and Maintenance | 6.4 Care and Cleaning Work
Interval
Component/Part
Every 6000 h*
(filament on)
Field emission gun (filament)*
Yearly or as required
(and after exchange of field emission gun)
§
Standard aperture
§
Anode aperture
§
Extractor aperture
§
Anode aluminum seal
§
Copper seal at gun head
Yearly or as required
Tip seal and tip O-ring of the pre-vacuum
pump
Yearly performance check
§
InLens SE detector
§
All optional detectors
* There is no warranty on field emission guns; manufacturers do not guarantee any lifetime.
Tab. 4: Schedule for the change of consumables
6.4 Care and Cleaning Work
All care and cleaning work not described here must only be performed by an authorized ZEISS ser-
vice representative.
NOTICE
Functional impairment due to dirt and moisture
Dirt, dust and moisture can impair the Microscope System’s functionality and can cause short-
circuits.
4
The ventilation slots must be unobstructed at all times.
4
Perform regular maintenance and cleaning according to the instructions in this document
and according to the instructions in the applicable documents.
4
Make sure that no cleaning liquid or moisture gets inside the Microscope System.
4
In case of damage, the affected parts of the Microscope System must be taken out of op-
eration.
6.4.1 Cleaning the Microscope
CAUTION
Isopropanol
Isopropanol is highly flammable and irritating to the eyes.
Vapors may cause drowsiness and dizziness.
4
Wear suitable gloves.
4
Keep away from sources of ignition.
4
Do not smoke.
4
In case of contact with the eyes, rinse immediately with plenty of water and seek medical
advice.
4
Avoid contact with skin.
4
Do not breathe vapor.
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000