ZEISS
9 Technical Data and Conformity | 9.2 Performance Data and Specifications | GeminiSEM 360
Parameter
Value
§
Black - LIVE
§
White - NEUTRAL
§
Green - PROTECTIVE EARTH
A fixed connection of the microscope re-
quires an additional mains switch in close
proximity to the microscope which complies
with IEC60947-1 and IEC60947-3 regula-
tions.
Power consumption
Max. 3600 VA, dependent on accessories
Max. current
16 A at 208–230 VAC
Addition building PE
The system must be connected to a building
earth point at all times.
Minimum cross section of an additional PE is
10 mm² Cu or 16 mm² Al
An exclusive grounding connection to earth
must be provided, i.e. the grounding terminal
must not be common to other electrical
equipment.
Circuit breaker (at house installation)
25 A (type K) according to IEC EN 60898
Circuit breaker type
Type K
Residual Current Device (RCD)
100 mA according to IEC EN 60898
Ampere interrupting capacity (AIC)
1500 A rms without external EMO box
10000 A rms with external EMO box
Max. leakage current (relevant for Residual
Current Device)
15 mA at 208–230 VAC (without further op-
tions which need 230 V supply)
Short Circuit Current Ratio (SCCR)
3.5 kA
Protection class
Class I
Overvoltage Category
II
Electron Optics
Parameter
Description
Acceleration volt-
age
Range:
20 V to 30 kV
Adjustment:
Continuously variable in 10 Volt steps
Probe current
20 nA high resolution configuration:
3 pA to 20 nA
100 nA high current configuration:
6 pA to 100 nA
Stability:
Better than 0.2 % per hour
Probe current selection
via seven apertures (six apertures in case of
100 nA configuration) with electromagnetic selection and precise
alignment
Optional OptiProbe
module for continuous probe current adjust-
ment
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000
135