9 Technical Data and Conformity | 9.4 Performance Data and Specifications | GeminiSEM 560
Detection System
Parameter
Description
Detection system
Detection system equipped for parallel detection and signal process-
ing of multiple detector channels:
Up to 16 detector channels (optional), 4 detector channels (default)
Up to 3 TV inputs (optional), 1 TV input (default)
Parallel detection, processing, and display of up to 4 channels
Any 2 detector signals can be mixed for enhanced image information.
TV Camera image and detector images can be displayed simultane-
ously.
InLens detectors
InLens SE detector:
Column-mounted scintillator detector with optically coupled photo-
multiplier for high efficiency surface sensitive InLens SE detection
EsB detector (optional):
Column-mounted high efficiency scintillator detector with optically
coupled photomultiplier for detection of energy and angle selective
backscattered electrons. Filtering grid adjustable from 0 V to −1.5 kV
for energy filtering
The InLens SE detector and the EsB detector can be used in parallel
Chamber detectors
aBSD/aBSD-LH detector (optional):
Pneumatically retractable 5 or 6 segment multi-mode solid state BSE
detector; enables materials contrast, crystal orientation, and topo-
graphic imaging. Enables in-situ 3D surface reconstruction and quanti-
tative 3D metrology.
VP-BSD detector (optional for NanoVP mode):
Pneumatically retractable 5 segment multi-mode solid state BSE de-
tector integrated into the beamsleeve aperture; enables materials con-
trast, crystal orientation, and topographic imaging. Enables in-situ 3D
surface reconstruction and quantitative 3D metrology.
aSTEM detector (optional):
Pneumatically retractable multi-mode annular Scanning Transmission
Electron Microscopy (aSTEM) detector, with 12x specimen holder; en-
ables bright field (BF), dark field (DF), oriented dark field (ODF), and
high angle annular dark field (HAADF) transmission imaging
C2D detector (optional):
Cascade Current Detector for detecting SE signal under variable pres-
sure with high efficiency.
Detector with floating amplifier electronics, for enhanced imaging in
Variable Pressure mode up to 500 Pa. Delivers detail at low kV for
specimens that demand higher pressures, e.g. polymers, bio speci-
mens, powders, and fibers. Enhanced signal-to-noise ration versus
VPSE detector.
CL detector (optional):
Cathodoluminiscence (CL) chamber detector
SCD detector (optional):
The specimen current detector (SCD) detects the current absorbed by
the specimen.
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000