ZEISS
5 Commissioning and First Operating Steps | 5.5 Modifying Gun Parameters
Info
After cathode replacement or after re-alignment of the electron optical column, OptiProbe has
to be calibrated.
Prerequisite
ü
Requires the licenses OPTIPROBE and HIGH VOLTAGE
ü
Particular hardware and the specimen current amplifier are installed.
Procedure
1. In the GeminiSEM Control panel, select the
Gun
tab.
2. Activate the
OptiProbe
checkbox.
3. Use the
I Probe
slider to set the desired probe current.
à
As soon as the probe current adjustment is finished, the
OptiProbe Status
readout
changes from
Busy
to
Ready
.
à
The probe current displayed in the
I Probe
readout corresponds to the actual probe cur-
rent (± 15 %).
4. To deactivate the function, deactivate the
OptiProbe
checkbox.
INFO:
If
OptiProbe
is deactivated, the
I Probe
readout continues to indicate the actual
probe current (± 15 %).
5.5.4 Measuring the Probe Current
Measuring the probe current using the Faraday cup ensures that the current displayed in the soft-
ware equals the incident probe current. The Faraday cup consists of a strongly absorbing material
with a cavity covered by a small aperture. If the beam is focused in this cavity, no secondary elec-
trons and no backscattered electrons leave the Faraday cup.
Info
The specimen current (absorbed current) is the fraction of the incident probe current (beam
current) that hits the specimen and is not emitted as secondary electrons, backscattered elec-
trons, or transmitted electrons. In case there is no other signal emission from the specimen,
the specimen current equals the incident probe current that hits the specimen.
Parts and Tools
Designation
Part no.
Faraday cup
348342-8055-000
Prerequisite
ü
You have successfully carried out the steps described in
Procedure
1. Load the Faraday cup into the specimen chamber. Refer to
2. Evacuate the specimen chamber.
3. Switch on the gun. Refer to
4. Switch on the EHT. Refer to
.
5. From the
Panel Configuration Bar
, select
Specimen Current Monitor
.
à
The
Specimen Current Monitor
window is displayed.
6. Activate the
Stage Bias
checkbox.
INFO:
If the
Stage Bias
checkbox is activated, a small negative bias voltage is applied to
the specimen stage. This enhances the sensitivity for touch alarm, especially for weakly con-
ducting specimens. In some special cases (e.g. very low EHT values), it may be necessary to
deactivate the Stage Bias option and set the stage potential close to zero.
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000
87