2 Safety | 2.5 Safety Devices and Interlocks
ZEISS
2.5.3 Circuit Breaker
The circuit breaker (F10) at the rear of the plinth will disconnect the mains power from the elec-
tronics in the plinth in case of an over-current (including the first pre-vacuum pump and PC). The
other circuit breaker (F11) disconnects mains power from the heaters and the second pre-vacuum
pump.
However, some other components will still be connected to mains power.
Fig. 3: Circuit breaker
The circuit breaker does not completely isolate the plinth from the mains power. This can only be
done using the Main Disconnect Device.
2.5.4 Emergency Off (EMO) Option External
The Emergency Off Circuit with the Main Switch is essential for compliance with the SEMI stan-
dard.
The Emergency Off Circuit is mounted at the wall near the mains power supply and connects the
mains power supply to the microscope and all its accessories.
Fig. 4: Emergency Off (
) Circuit
With the optional Emergency Off Circuit, the microscope is equipped with the following additional
safety equipment.
22
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000