7 Troubleshooting | 7.4 Power Circuit
ZEISS
à
The
Probe Current Calibration
window is displayed.
6. Activate the
Spot
checkbox.
7. Click
Cal I Probe
.
8. To confirm, click
Yes
.
9. To store the calibration, click
Save
.
10. Deactivate the
Spot
checkbox.
7.3.3 Calibrating the Stigmator
If the image shifts when you change the stigmator, then you need to set the stigmator balance.
The calibration needs to be done for only one single probe current, e.g. 30kV:10pA.
The SEM stigmator calibration is an alignment process that enables you to remove effects caused
by astigmatism. Astigmatism causes a circular beam to become elliptical. As a result, the focused
live image appears to be blurred in one direction (line focus).
Procedure
1. From the
Panel Configuration Bar
, select
SEM Stig Calibration
.
à
The
SEM Stig Calibration
panel is displayed.
2. In the
Wobble
section, select
Stig X
or
Stig Y
.
INFO:
If the image appears to be blurred in X direction, then select
Stig X
.
3. Set the stigmator balance period via the
Stig Bal Period
scroll bar.
INFO:
A good starting value is 1000. For higher scan speeds (slower), increase the value.
For lower scan speeds (faster), reduce the value.
4. Set the stigmator balance amplitude via the
Stig Bal Amplitude
scroll bar.
INFO:
A good starting value is 100 %. For bigger magnifications, decrease the value.
5. Adjust the stigmator via the
Stig Centre
navigation box.
6. If the focused live image is still blurred in one direction or the image still shifts when you
change the stigmator, then repeat the procedure.
7.4 Power Circuit
7.4.1 Checking the Position of the Circuit Breakers
NOTICE
Persisting electrical problems
Tripped circuit breakers may be a hint for an electrical problem in the microscope.
4
If a circuit breaker keeps tripping, de-energize the microscope completely and contact your
ZEISS service representative for assistance.
No.
Value
Circuit
F1
10 A
Power supply unit
F2
10 A
RED (X20) & Green (X26) accessories
F3
10 A
Pre-Vacuum pump (X25) & Yellow (X22) ac-
cessories
Aux 1-2 + 2 Spare labeled FIB,GIS (Crossbeam
relevant only)
F4
10 A
Internal heaters
F5
10 A
Circuit surge protection off
124
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000