9 Technical Data and Conformity | 9.4 Performance Data and Specifications | GeminiSEM 560
ZEISS
Parameter
Description
SE detector:
Everhart-Thornley SE detector with optically coupled photomultiplier;
collector bias adjustable from −250 to +400 V
VPSE detector (optional):
Light detector for detecting SE signal under variable pressure (VP)
with high speed.
YAG detector (optional):
Manual retractable, high speed 1 sector scintillator based BSE detec-
tor.
Chamber camera
Color CCD camera
with white-light illumination and IR illumination.
Specimen current
monitor
6 range auto ranging for precise current measurement in the range of
1 pA to 10 μA
Standard VP
(option)
Parameter
Description
Vacuum range
Vacuum range from 5 Pa to 60 Pa, adjustable in steps of 1 Pa
VP detection
At least one of the following detectors must be selected in the stan-
dard VP option:
§
High speed Variable Pressure Secondary Electron Detector (VPSE)
§
High efficiency Cascade Current Detector (C2D)
§
Backscatter Electron Detector (aBSD/aBSD-LH)
NanoVP (option)
Parameter
Description
Beamsleeve
Retractable beamsleeve aperture below the objective lens to shorten
the beam gas path length and to enable InLens detection in VP mode.
Vacuum range
Vacuum range from 5 Pa to 500 Pa, adjustable in steps of 1 Pa
VP detection
Additional to the detectors selected in standard VP option, the VP-
BSD detector is available in NanoVP mode.
150
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000