ZEISS
5 Commissioning and First Operating Steps | 5.8 Working with Variable Pressure
5. Click
Next
.
à
The system is vented.
à
The beamsleeve is inserted, but not fully so
that you have access to the beamsleeve
aperture inside the specimen chamber.
6. Open the chamber door.
If there is already a beamsleeve aperture in, use plastic tweezers to carefully take it out.
Return the aperture to the beamsleeve aperture kit.
It is not possible to insert another aperture at the same time.
NOTICE
Be careful not to touch the objective lens or parts inside the specimen
chamber. Wear gloves and forearm protectors.
7. Use plastic tweezers to carefully take the cor-
rect beamsleeve aperture (350 µm or 800 µm)
out of the beamsleeve aperture kit.
Carefully insert the beamsleeve aperture.
8. Close the chamber door.
9. In the
Beamsleeve Configuration
wizard, select the beamsleeve aperture you just have in-
stalled.
10. Click
Finish
.
à
The beamsleeve is retracted and the system is pumping.
11. Remove the stopper from the slide.
5.8.2.2 Aligning the Beamsleeve Aperture
After you have installed a beamsleeve aperture you should center it with the
Beamsleeve Align-
ment
window.
The procedure consists of the following steps:
§
Inserting the Beamsleeve Aperture [
§
Measuring the Size of the Beamsleeve Aperture [
§
Centering the Beamsleeve Aperture [
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000
107