ZEISS
5 Commissioning and First Operating Steps | 5.10 Performing an Emergency Shutdown
à
All power is cut off from the microscope.
6. To disconnect the power connection at the facility installation, remove the power plug
(male connector) from the socket (female connector).
7. To secure the microscope against accidental re-activation, refer to
.
5.9.3 Performing a Lockout/Tagout
The Lockout/Tagout (LOTO) procedure defines the process by which the microscope is isolated
from potential hazards associated with the unexpected release of hazardous energy. This proce-
dure needs to be performed when service, maintenance work, or unscheduled repairs can cause
exposure to any form of hazardous energy.
Info
The customer is responsible for providing instructions on how to operate the energy-isolating
devices properly.
Prerequisite
ü
The microscope has been de-energized.
Procedure
1. Verify that the microscope has been de-energized properly.
2. To secure the microscope against accidental re-activation, lock the energy-isolating devices.
5.10 Performing an Emergency Shutdown
NOTICE
Components in the high voltage circuitry
When the microscope, especially the gun, is fully on, an abrupt shutdown of all electrical sup-
plies may damage some components in the high voltage circuitry, mainly the cathode.
4
Use the emergency off only in an emergency situation with personnel injury.
Procedure
1. Depending on whether the EMO option is installed on your microscope or not, operate the
respective control element.
If the EMO option is installed, press the
EMO
button on the top of the plinth.
If the EMO option is not installed, set the
MAIN
switch at the rear side of the plinth to the
OFF position.
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000
113