Setting of Z focus factors and parfocality
The matching of focus positions (parfocality) and the adaptation of the focus speed (Z-focus factors) to
the objective magnification are normally only performed when the microscope is used for the first time
or after the change of objectives in the nosepiece.
The parfocality values are automatically saved during setting of the Z focus factors. To correctly save the
parfocality values, the objective with the highest magnification must be used to focus on a specimen
before the Set mode is entered.
•
Place a specimen on the stage. Swing in dry objective with the highest magnification.
•
Focus on the specimen via the fine adjustment.
•
Activate Set mode by pressing key
Set LM
(for more than 2 seconds).
•
Press one of the two
FOCUS
keys.
Change to focus programming is made, and
SET FOC SPEED 6
is displayed in the upper line of the LCD
display. The number
6
stands for the focus factor currently set. The set nosepiece position and the
objectives in the nosepiece are shown in the lower LCD line.
•
Activation of one of the two
FOCUS
keys permits the focus factor to be changed individually for the
swung-in objective.
Key FOCUS :
the drive moves faster
Key FOCUS :
the drive moves slower
•
Then focus on the specimen again and swing in the objective with the next lowest magnification.
While the nosepiece is being moved to the next position, the focus position and focus speed values set
for the previous objective are saved in the Set mode.
•
Use the current objective to set the focus speed, focus exactly on the specimen and move the
nosepiece to the next position to store the settings.
•
Repeat the above settings for all objectives. The last settings should always be made with the oil
immersion objectives.
•
Finally, the nosepiece must be turned by one further position to permit the values to be saved.
•
To save the entire setting, exit the Set mode by briefly pressing the
Set LM
key.
Resetting of Z focus factors and parfocality to default values
•
Activate Set mode.
•
Press and hold FOCUS and FOCUS keys simultaneously for approx. 1 s until you hear a beep.
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...