![background image](http://html1.mh-extra.com/html/zeiss/axiovert-200/axiovert-200_operating-manual_3384802051.webp)
START-UP
Axiovert 200
HAL 100 W halogen illuminator
Carl Zeiss
B 40-080 e 03/01
2-21
2.12.2
Coarse alignment of halogen
illuminator
•
Loosen the clamping screw (2-27/
2
) and
remove the halogen illuminator from the carrier
for transmitted-light illumination.
•
Switch on the microscope.
•
Direct light beam against a projection area
(wall) approx. 3 m away.
Make sure not to look into the light
exit opening of the illuminator.
•
Use SW 3 ball-headed screwdriver to set
adjusting screw (2-26/
3
) in such a way that
both images of the lamp coil on the projection
area are defined as clearly as possible.
•
Then set adjusting screws (2-26/
4
and
5
) in
such a way that the lamp coils of one image
exactly fill the gaps of the reflector image
(2-26/
1
).
2.12.3
Attachment of halogen illuminator
•
If required, remove cover from the lamp mount
(2-27/
2
) in the carrier for transmitted-light
illumination.
•
Insert dovetail (2-27/
2
) of lamp housing (2-26/
3
)
in carrier (2-27/
2
) and use SW 3 ball-headed
screwdriver to tighten clamping screw (2-27/
1
).
•
Connect 3-pin lamp plug (2-27/
4
) to 3-pin 12 V
100 W socket (2-27/
5
) on the instrument rear.
Fig. 2-26
Alignment of halogen illuminator
Fig. 2-27
Attachment of halogen illuminator
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...