START-UP
Carl Zeiss
HAL 100 W halogen illuminator
Axiovert 200
2-20
B 40-080 e 03/01
2.12
HAL 100 halogen illuminator
2.12.1
Change / attachment of the HAL
100 halogen lamp
•
Switch off the microscope, remove connector
(2-27/
4
) from the 12 V / 100 W socket (2-27/
5
)
and allow the illuminator to cool down for
approx. 15 minutes.
•
Loosen screw (2-25/
5
) until the lamp housing
can be removed in an upward direction.
•
Press both spring levers (2-25/
3
) downwards
and pull out the old halogen lamp in an upward
direction.
•
Press both spring levers downwards, insert new
lamp (2-25/
2
) with protection cap (2-25/
1
) in
lamp base (2-25/
4
), let go the spring lever and
pull off the protection cap.
•
Press spring lever briefly downwards again to
center the lamp.
•
Attach lamp housing again and tighten
clamping screw (2-25/
5
).
☞
After the lamp change, the halogen
illuminator must be realigned.
Fig. 2-25
Changing the halogen lamp.