CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Carl Zeiss
Maintenance
Axiovert 200
4-4
B 40-080 e 03/01
4.2
Maintenance
4.2.1
Performing checks
The following checks should be performed regularly every six months.
General
•
Check line cables and plugs for defects.
•
If any damage is visible, switch off the instrument. Have defects repaired by Zeiss service staff.
•
Make sure that the maximum operation time of the halogen and HBO / XBO lamps is not exceeded
(once a week).
Illumination
•
Check the correct setting of the halogen and HBO / XBO lamps.
•
Check contacts of the electrical connection points of the illuminators.
Optics
•
Visual checking of cleanliness of objectives and eyepieces
4.2.2
Changing the fuses on the microscope
The fuses for the instrument input voltage are
positioned on the rear of the microscope stand.
The fuse compartment, which contains two
T4.0
A/H 250 V 5x20 mm
fuses, is combined with the
instrument socket.
If fuses blow, the cause must be found out first
and any technical defect must be efficiently
eliminated.
Always pull the power plug before
changing the fuses.
•
Disconnect the instrument from the line. Pull
out fuse compartment (4-1/
1
) in forward
direction.
•
Remove fuses from the holder and insert new fuses.
•
Insert fuse holder back in the fuse compartment (4-1/
2
) until stop. Plug in power cords.
Fig. 4-1
Changing the fuses
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...