CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axiovert 200
Care
Carl Zeiss
B 40-080 e 03/01
4-3
4
CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
4.1
Care
Care of the instruments is limited to the following operations:
•
Cover the instrument with the dust cover after every use.
•
Do not install instrument in a damp room.
•
Cover open tubes with dust protection caps.
•
Remove dust and dirt from visible optical surfaces using a brush, airblower, Q-tips, optics cleaning
paper or a cotton cloth.
•
Remove water-soluble dirt (coffee, Coke, etc.) after breathing on it and wiping it off with a dust-free
cotton cloth or a moistened cloth. A mild cleaning agent can also be added to the water.
•
Remove stubborn, oily or greasy dirt (immersion oils, finger prints) with a wad of cotton or a dust-free
cotton cloth dipped in the optics cleaning mixture L.
This cleaning mixture is produced of 90 Vol% benzoline and 10 Vol% isopropanol (IPA). The various
components are also known under the following synonyms:
benzoline:
medical alcohol, petroleum ether
Isopropanol:
2-propanol, dimethyl carbinol, 2-hydroxypropane
Clean the optical surface by moving in circles starting in the middle. Slight pressure should be exerted on
the optics during cleaning.
When using the Axiovert 200 in humid climatic zones, proceed as follows:
•
Store the microscope in bright, dry and well ventilated rooms with a humidity of less than 65%; store
particularly sensitive components and accessories, such as objectives and eyepieces, in a dry closet.
•
When the equipment is stored in closed cases for a longer period of time, the growth of fungus can
be avoided by including cloths soaked in fungicide in the cases.
The risk of growth of fungus on opto-mechanical instruments always exists in the following
conditions:
−
relative humidity of more than 75% and temperatures b15° C and +35° C for
more than three days.
−
installation in dark rooms without air ventilation, and
−
dust deposits and fingerprints on optical surfaces.
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...