START-UP
Axiovert 200
Illuminator N HBO 103
Carl Zeiss
B 40-080 e 03/01
2-25
2.13.2
Coarse alignment of illuminator N
HBO 103
The following settings for adjustment of the
N HBO 103 illuminator are performed in two steps:
the coarse adjustment and the subsequent fine
adjustment using the adjustment aid.
During coarse adjustment, UV
protection goggles must be worn to
protect the eyes from UV radiation.
•
Remove illuminator from the microscope stand
by loosening the clamping screw (2-35/
2
).
•
Switch on the entirely connected illuminator
N
HBO
103 (also see page 2-26) via the
ebq 100 dc power supply (2-24/
2
) and allow it
to heat to operating temperature. Make sure not to look into the light exit opening of the
illuminator.
•
Direct light beam against a projection area (wall) approx. 3 m away.
•
Use knurled knob (2-32/
6
) to focus on the brighter focal spot via collector adjustment.
•
Use adjusting screw (2-32/
3
) and SW 3 ball-headed screwdriver to locate and focus on the slightly
darker, reflected focal spot.
•
Use adjusting screws (2-32/
4
and
5
) and SW 3 ball-headed screwdriver to set the height and side of
the brighter focal spot in accordance with the focal spot image (2-34).
Fig. 2-33
Focal spots of N HBO 103 before coarse
adjustment
Fig. 2-34
Focal spots of N HBO 103 after
coarse adjustment
Fig. 2-32
Coarse alignment of mercury vapor
short-arc lamp
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...