![background image](http://html1.mh-extra.com/html/zeiss/axiovert-200/axiovert-200_operating-manual_3384802053.webp)
START-UP
Axiovert 200
Illuminator N HBO 103
Carl Zeiss
B 40-080 e 03/01
2-23
2.13
Illuminator N HBO 103
2.13.1
Attachment / change of the HBO 103 W/2 mercury pressure short-arc lamp
The illuminator N HBO 103 and the HBO 103 W/2 mercury pressure short-arc lamp are supplied to the
customer in separate packaging for safety reasons.
Therefore, insertion of the HBO 103 W/2 into the lamp housing is the first step in the start-up of the
illuminator.
Before opening the lamp housing, make sure that no connection to the electrical line is
available.
The HBO 203 W/2 lamps may only be changed after they have cooled down. Allow the
N HBO 103 microscope illuminator to cool down for approx. 15 minutes to avoid the risk of
explosion and burns.
The HBO 103 W/2 lamp may only be removed from the packaging and inserted in the
N HBO 103 illuminator if a protection mask and safety gloves are worn.
All electrical clamping connections must be made carefully. Pronounced heat during operation
may result in loose contacts.
For start-up or after expiry of the average
operation time of 300 h, the HBO 103 W/2
mercury vapor short-arc lamp must be inserted or
exchanged.
The operating time of the lamp can be read from
the counter (2-29/
1
) of the ebq 100 dc power
supply.
Follow the following steps when inserting /
changing the HBO 103 W/2 lamp:
•
Pull off or unscrew line cable (near 2-29/
2
) and
N HBO 103 connector (near 2-29/
3
) from the
ebq 100 dc power supply.
•
Wear protection mask and safety gloves.
Fig. 2-29
Power supply ebq 100 dc
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...