START-UP
Carl Zeiss
Attachment of condensers
Axiovert 200
2-14
B 40-080 e 03/01
•
Insert condenser adapter (2-14/
1
) into the condenser carrier on the carrier for transmitted-light
illumination with its dovetail pointing upwards. Make sure that the orientation pin of the condenser is
positioned at the front and exactly engages into the guiding groove of the condenser carrier.
•
Fix condenser adapter in position with clamping screw (2-14 /
2
).
•
Then insert the required condenser into the condenser adapter via the dovetail, making sure that the
orientation is correct, and fix it into position with clamping screw (2-14/
3
).
2.6.3
Changing the DIC prism in the
condenser turret
•
To change a DIC prism, remove the condenser
and place it upside down on a sturdy support to
make the underside accessible.
•
Remove the plastic cover (2-15/
1
) from the
assembly opening (2-15/
3
).
•
Position the turret disk containing the DIC prism
to be exchanged in the assembly opening and
hold it on the knurled ring.
•
Use mounting device from the tool set to
unscrew retainer ring (2-15/
2
).
•
Turn condenser upside down and allow DIC
prism (2-15/
4
) to slide out onto a soft surface.
Installation of the DIC prism is made in reverse
order:
•
Carefully insert new DIC prism into the
assembly opening with the labeling pointing
upwards. If required, use tweezers to hold the
DIC prism carefully on its outer ring. Take
special care of the correct orientation of the
DIC prism in the mount (groove of the DIC
prism must engage in the pinion of the mount).
•
Carefully insert retainer ring again and screw it tight using the mounting device.
•
Use plastic cover to close the assembly opening again.
•
Make sure that the knurled ring of the turret disk is labelled correctly.
•
Turn the condenser around and insert it in the carrier for transmitted-light illumination.
Fig. 2-15
Changing the DIC prism
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...