CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axiovert 200
Maintenance
Carl Zeiss
B 40-080 e 03/01
4-5
4.2.3
Changing the fuses on the ebq 100 dc
power supply
The fuse holder for the fuses F1 and F2 is
positioned on the rear of the power supply. It is
combined with the instrument socket and contains
two
T 3, 15A
fuses.
If fuses blow, the cause must be found out first
and any technical defect must be efficiently
eliminated.
Always pull the power plug before
changing the fuses.
•
Disconnect the instrument from the line.
•
Pull fuse holder (4-2/
1
) out of the fuse
compartment (4-2/
2
) in forward direction.
•
Replace defective fuse.
•
Then push fuse holder back into the fuse
compartment until click-stop.
•
Plug in power cords.
Fig. 4-2
Changing the fuses on the power
supply