START-UP
Carl Zeiss
HAL 100 W halogen illuminator
Axiovert 200
2-22
B 40-080 e 03/01
2.12.4
Fine alignment of halogen
illuminator
Fine alignment requires the diffusion disk to be
removed:
•
Loosen clamping screw of HAL 100 W (2-28/
1
)
and remove the illuminator from the carrier for
transmitted-light illumination.
•
Turn diffusion disk (2-28/
2
) out of the carrier
manually (anti-clockwise). Use the projections
(2-28/
3
) on the mount of the diffusion disk to
hold the component.
•
Attach HAL 100 W and tighten the clamping
screw.
•
Remove any swung-in filter from the beam
path.
•
Use objective
≤
40x to focus on the specimen and search for an empty object spot.
•
Remove eyepiece and use adjusting screws (2-26/
4
and
5
) to center the lamp coil and its reflected
image in the pupil image.
•
Use adjusting screws (2-26/
3
) to optimize the homogeneous illumination of the pupil image.
•
Remove HAL 100 W after conclusion of the alignment.
•
Screw diffusion disk into the carrier again manually.
•
Attach HAL 100 W and swing in available filters again.
Fig. 2-28
Changing the diffusion disk
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Page 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...