OPERATION
Carl Zeiss
Axiovert 200 M
Axiovert 200
3-20
B 40-080 e 03/01
Motorized condensers
(3-13/
16
)
Motorized condenser, equipped with:
−
6-position turret for:
brightfield: H
phase contrast: Ph1, Ph2, Ph3 with centerable
stops
interference contrast: DIC I, DIC II
−
If a motorized condenser is recognized by the
Sub-bus of the stand, the current condenser
turret position automatically appears in the
lower line of the LCD display. The condenser
display is always available.
−
Motorized adjustment of the turret by pressing
key
Rev
(forward and backward, 3-15/
1
)
on the right side of the condenser.
−
Motorized adjustment of the aperture
diaphragm by pressing key
A
(open and close,
3-15/
2
) on the left side of the condenser. If a
phase stop is contained in the beam path, the
aperture diaphragm always fully opens
automatically (NA = 0.55).
−
The set turret position appears in the lower line
of the LCD display (e.g.
Cond. DICII
).
Cond.
NA: 0.55
is displayed in a phase stop position,
and the key pair
A
is without function.
☞
The motorized condenser can only be used with the motorized stand, or with a manual stand
which is equipped with an interface PC-board.
1x OPTOVAR key for the motorized change of Optovar settings
(3-13/
12
)
−
Key for the selection of the Optovar magnification; switching sequence 1x, 1.6x, 2.5x
−
Key pressed briefly:
current position is shown in the LCD display for 4 s (or until the next press
of the key);
without LCD display, the position is approached imediately
key pressed again:
target position is displayed and approached
−
In the LCD display, the selected Optovar magnification and the overall magnification are shown.
Fig. 3-15
Condenser 0.55, 6-position
H, Ph1, Ph2, Ph3, DIC, DIC
Summary of Contents for Axiovert 200
Page 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Page 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Page 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Page 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...